Function not available for verified balances.
Give balance factory number
Format: NB CR LF
Accessible responses:
NB_A_”x” CR LF
- command understood, balance factory number is
given in response
NB_I CR LF
- command understood but not accessible at this
moment
xbalance factory number (inbetween inverted comas)
Example
Command:
NB CR LF give balance factory number
Response:
NB_A_”1234567” balance factory number”1234567”
Lock balance keyboard
Format: K1 CR LF
Accessible responses:
K1_I CR LF
- command understood but not accessible at this
moment
K1_OK CR LF
- command completed
Command locks balance keyboard (motion sensors, touch panel) until it gets
unlocked either by K0 command or by switching the balance off.
Unlock balance keyboard
Format: K0 CR LF
Accessible responses:
K0_I CR LF
- command understood but not accessible at this
moment
K0_OK CR LF
- command completed
Start BEEP signal
The command activates BEEP signal for a specified time limit.
Format: BP_TIME <CR><LF>
Accessible responses:
BP_OK <CR><LF>
command completed, BEEP signal activated
BP_E” <CR><LF>
incorrect format or lack of the parameter
BP_I <CR><LF>
command understood but not accessible at this
moment
TIME parameter, decimal value, determined time limit for the signal, given
in [ms]. Recommended range <50
÷ 5000>.