Function not available for verified balances.
Give balance factory number
Format: NB CR LF
Accessible responses:
- command understood, balance factory number is
given in response
- command understood but not accessible at this
moment
x – balance factory number (inbetween inverted comas)
Example
Command:
NB CR LF – give balance factory number
Response:
NB_A_”1234567” – balance factory number – ”1234567”
Lock balance keyboard
Format: K1 CR LF
Accessible responses:
- command understood but not accessible at this
moment
Command locks balance keyboard (motion sensors, touch panel) until it gets
unlocked either by K0 command or by switching the balance off.
Unlock balance keyboard
Format: K0 CR LF
Accessible responses:
- command understood but not accessible at this
moment
Start BEEP signal
The command activates BEEP signal for a specified time limit.
Format: BP_TIME <CR><LF>
Accessible responses:
– command completed, BEEP signal activated
– incorrect format or lack of the parameter
– command understood but not accessible at this
moment
TIME – parameter, decimal value, determined time limit for the signal, given
in [ms]. Recommended range <50
÷ 5000>.