Radwag MA 200.3Y Moisture Analyzer Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Radwag MA 200.3Y Moisture Analyzer. RADWAG MA 200.3Y Moisture Analyzer User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 170
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 0
User manual no.:
IMMU-10-03-09-14-ENG
MOISTURE ANALYZER
MA 3Y
SERIES
BALANCES AND SCALES
RADWAG 26 600 Radom, 28 Bracka Street, POLAND
Phone: +48 (0-48) 38 48 800, fax. +48 (0-48) 385 00 10
export@radwag.com
www.radwag.com
Vista de página 0
1 2 3 4 5 6 ... 169 170

Resumo do Conteúdo

Página 1 - MOISTURE ANALYZER

User manual no.: IMMU-10-03-09-14-ENG MOISTURE ANALYZER MA 3Y SERIES BALANCES AND SCALES RADWAG 26 – 600 Radom, 28 B

Página 2

- 10 - Corrosion Substances that release aggressive vapours (e.g. acids) during the heating process. In this case, it is recommended to work with

Página 3

- 100 - Caution: It is possible to use an option for searching a report. List of parameters in a report from drying process: 1. Start date 2.

Página 4

- 101 - The bottom bar contains the following soft keys (options): Return to displaying the complete chart Reducing the screen to the previous vi

Página 5

- 102 - An instance of a chart is presented below. The bottom bar contains the following soft keys (options: Return to displaying the complete

Página 6

- 103 - 19.8. Minimal sample weight Minimal sample weight database contains data for declared method and for a given balance minimal sample weight

Página 7 - 1. TECHNICAL SPECIFICATION

- 104 - ranging from 0.0001 g to 10.0000 g (<TARE> button in use) 3 50.0000 g 2.5000 g Minimal sample weight for all net masses being weighed i

Página 8

- 105 - • Press field of a desired report record, unless it is visible on the top of the list, wind the list forward using navigating keys • Report

Página 9

- 106 - 19.12. Printouts Database of Printouts contains all saved NON-STANDARD printouts. Each non-standard printout features a name, code and a s

Página 10

- 107 - 1 – enlarging the editing field (7), recommended while using an external computer keyboard connected to balance’s USB port 2 – cancellin

Página 11

- 108 - Example of Printout no. 2 – printout template from a file A file with printout template has to have an extension: *.txt or *.lb and contain

Página 12

- 109 - 19.14.1. Export database of weighing records to file All carried out measurements are saved and stored in a database. The moisture an

Página 13

- 11 - At the initial stage of measurement, the device precisely determines the mass of an object placed on instrument’s weighing pan. As the mass

Página 14 - 3. TRANSPORT AND STORAGE

- 110 - • Press < Export weighing database to a file > button, the software automatically starts exporting the database to a file, • On comp

Página 15

- 111 - Upon confirmation the software deletes data and displays window with number of deleted records: Upon confirmation the software re

Página 16

- 112 - Upon confirmation of the entered deadline date all weighing and reports dating on a given period of time will be deleted. Number of dele

Página 17

- 113 - 20. COMMUNICATION Menu Communication is located in a menu group Parameters. It is accessed by pressing < Setup > quick access key

Página 18

- 114 - Return to weighing mode and restart the device. 20.3. WiFi settings For balance equipped with WiFi module the top toolbar of the main d

Página 19

- 115 - Icon with a padlock, placed by a given network name requires password for the network. Press <Refresh> icon to search for available n

Página 20

- 116 - Procedure of setting a port for “TCP” protocol: − Enter group of parameters < Communication>, − Select: „ Tcp / Port” which

Página 21 - 5. STARTUP

- 117 - 21. PERIPHERAL DEVICES Menu PERIPHERAL DEVICES is located in menu group Parameters. It is accessed by pressing Setup key or < Setup

Página 22 - 6. KEYBOARD – FUNCTION KEYS

- 118 - 21.2. Printer Moisture analyzer’s submenu < Printer > enables: − Setting communication port with a printer, − Defining printer’

Página 23

- 119 - • control codes – 1B7412 Default values for each of printout standards: Product printout template {50}

Página 24 - 7. SOFTWARE STRUCTURE

- 12 - case of any problems it is recommended to contact RADWAG service or a distributor. It must be always remembered to use the device in accordanc

Página 25 - 8. SOFTWARE MAIN WINDOW

- 120 - communicating the moisture analyzer with a barcode scanner is provided in APPENDIX F of this user manual.. The moisture analyzer enables com

Página 26

- 121 - • Enter the database of operators and edit desired operator’s record, go to < Card number> option, • On entering editing mode in <

Página 27

- 122 - 22. INPUTS / OUTPUTS Application of inputs: The set of inputs is used to control moisture analyzer’s operation. Each of 4 inputs is co

Página 28 - 9. LOGGING

- 123 - Application of outputs: The set of outputs is used to SIGNAL the status of measurement result. Change of logic status of the output syste

Página 29

- 124 - 23. OTHER PARAMETERS This menu contains global parameters on moisture analyzer operation, such as: language, date-time, beep sound, disp

Página 30 - 10. OPERATING USER MENU

- 125 - 23.3. “Beep” sound Procedure: Enter submenu < Others>, press < Beep> and set one of available options. No - Beep sound

Página 31 - 11. WEIGHING

- 126 - On entering the option enter a desired value [0% - 100%]. Upon entering a particular value the brightness changes respectively and the progra

Página 32

- 127 - Caution: Means of leveling the moisture analyzer is described in point 11.3 of this user manual. 23.9. Drying temperature reducing This

Página 33

- 128 - analyzer MA 3Y series it is not necessary to determine the drying parameters for the products to be dried, but simply use the selected and pr

Página 34

- 129 - 23.14. Start logo Caution: This parameter is available only for authorized operator of a moisture analyzer. The parameter enables changing

Página 35

- 13 - Please read the user manual carefully before instrument’s plugging to mains and startup, even if the user is experienced with this type of moi

Página 36 - • AUTOTARE

- 130 - The process of updating the weighing module takes places analogously to above procedure, but the required extension of the update file is: *.

Página 37 - • NON-STANDARD PRINTOUT

- 131 - 26. MOVIES Option activating instructional movie explaining how to operate balance functions. To view the movie go to <Movies> parame

Página 38 - • MODE

- 132 - To play the movie the user has to: 1. Save the movie to the balance memory (see instruction above) 2. Enter <Movies> parameter grou

Página 39

- 133 - 27. COMMUNICATION PROTOCOL General information A. A character based communication protocol moisture analyzer-terminal is designed for est

Página 40

- 134 - K1 Lock balance keyboard K0 Unlock balance keyboard BP Start BEEP signal OMI List available working modes OMS Set working mode OMG Give

Página 41

- 135 - 27.2. Responses format for commands sent from computer level On receipt of a command, the terminal sends a response in one of the following

Página 42 - 12. ADJUSTMENT

- 136 - Tare the moisture analyzer Format: T CR LF Accessible responses: T_A CR LF T_D CR LF - command understood and in progress - command car

Página 43

- 137 - Send stable measurement result in basic measuring unit Format: S CR LF Accessible responses: S_A CR LF S_E CR LF - command understood an

Página 44

- 138 - Send stable measurement result in current weighing unit Format: SU CR LF Accessible responses: SU_A CR LF SU_E CR LF - command unders

Página 45

- 139 - Accessible responses: C1_I CR LF - command understood but not accessible at this moment C1_A CR LF MASS FRAME - command understood and in

Página 46

- 14 - 3. TRANSPORT AND STORAGE 3.1. Delivery check Please check the packaging immediately upon delivery and check instrument’s condition directl

Página 47

- 140 - ES CR LF - command not recognized (incorrect mass format) Set high checkweighing threshold HI Format: UH_XXXXX CR LF, where: _ - spa

Página 48

- 141 - User log out Format: LOGOUT CR LF Accessible responses: LOGOUT OK CR LF - command understood, user is logged out ES CR LF - command not

Página 49

- 142 - Format: BP_TIME <CR><LF> Accessible responses: BP_OK <CR><LF> – command completed, BEEP signal activated BP_E” <CR

Página 50 - Hint:

- 143 - Response: OMI <CR><LF> – available working modes are given 2_” Parts counting” <CR><LF> number + name

Página 51

- 144 - n – parameter, decimal value specifying working mode number. See OMI command description for detailed information. Example Command: OMG<

Página 52 - 13. USERS

- 145 - Example: _ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF – a printout generated by a moisture analyzer on pressing key, according to settings of &l

Página 53 - 14. ACCESS LEVEL

- 146 - Cable: moisture analyzer - printer KAFKA Cable: moisture analyzer – printer CITIZEN, EPSON Cable: IN / OUT 30. ERROR MESSAGES Err2

Página 54

- 147 - 31. MOISTURE ANALYZER MAINTENANCE ACTIVITIES This section of the user manual describes procedures of maintaining a moisture analyzer MA 3Y

Página 55 - 15. PROFILES

- 148 - 31.2. Temperature sensor cleaning In order to ensure correct temperature measurements during testing procedure, pay special attention to c

Página 56

- 149 - 31.3. Replacing the fuses If after plugging to mains, a signal diode on moisture analyzer’s display does not light up (no display reaction)

Página 57

- 15 - 4.2. Unpacking Carefully remove the moisture analyzer from its packaging, remove the plastic and foil transport protective elements. Gently

Página 58

- 150 - 31.4. Troubleshooting Problem: no reaction to pressing main switch on/off key (no display reaction). Probable cause: • No voltage in mai

Página 59

- 151 - 32. ADDITIONAL EQUIPMENT Type Name P0136 RS232 cable for KAFKA printer P0151 RS232 cable for EPSON printer KAFKA Thermal printer EPSON

Página 60

- 152 - 33. APPENDIX A – Variables for printouts 33.1. List of variables Caution: Each of defined variables must be contained in braces:

Página 61

- 153 - {30} Gross value {31} Weighing platform no. {32} Factory no. {33} Balance reading unit {34} Range {35} Parts counting: Mass of a st

Página 62

- 154 - {75} User: Name {76} User: Code {77} User: Access level {80} Packaging: Name {81} Packaging: Code {82} Packaging: Mass {85} Client: Name {8

Página 63

- 155 - {121} 2) Average Tare Report: Result {122} 2) Average Tare Report: Standard deviation {123} 2) Average Tare Report: 0,25T1 {124} 2) Average T

Página 64 -  Non-standard printout

- 156 - {182} 2) Report from formulation: Measurements {188} 2) Ingredient: Target mass {189} 2) Ingredient: Difference {190} 2) Comparator: Repo

Página 65

- 157 - {223} 2) Mass control: Average {224} 2) Mass control: Standard deviation {225} 2) Mass control: Standard quantity {226} 2) Mass control: Meas

Página 66

- 158 - {268} 2) Density: Density of steel standard {269} 2) Density: Density of aluminum standard {275} Ambient conditions report: Date and time {

Página 67

- 159 - {317} 2) Pipettes: Type {318} 2) Pipettes: Volume type {319} 2) Pipettes: Tested volumes {325} 2) Pipette calibration report: Serial no. {3

Página 68

- 16 - When assembling the pan supporter (bracket) pay special attention to the correct positioning of its mandrel. The mandrel features a cut for i

Página 69

- 160 - {361} 2) Sample number {365} 2) Weighing B: Interval {366} 2) Weighing B: Difference {367} 2) Weighing B: Difference % {368} 2) Weighing B:

Página 70

- 161 - 33.2. Variables formatting Moisture analyzer MA 3Y series enables random formatting of numerical variables, text variables and date which a

Página 71

- 162 - Sign Description Example T Formatting mass and quantities derivatives of mass in EAN128 code {7:T6.3} – Net mass in a form of EAN128 code

Página 72 - 18. DRYING PROCESS

- 163 - 34. APPENDIX D – Settings of CITIZEN printer Baud rate 9600b/sec Parity control none Bits 8bit

Página 73

- 164 - 36. APPENDIX F – Communication with barcode scanners 1. For communication with barcode scanners RADWAG moisture analyzers MA 3Y series use

Página 74

- 165 - 37. APPENDIX G – Menu structure User adjustment Description Value External adjustment Function User adjustment Function Adjustmen

Página 75

- 166 - Databases − Products − Weighing records − Clients − Drying programs − Reports from drying process (Drying reports) − Ambient conditio

Página 76

- 167 - Peripheral devices Description Value Computer - » Port COM 1 » Address 1 » Continuous transmission No / Yes » Printout template

Página 77

- 168 - » Output 1 Selection » Output 2 Selection » Output 3 Selection » Output 4 Selection Access level Description Value Anonymou

Página 78

- 169 - Date and time Settings Beep Selection Touch screen display calibration Function Level control Selection Autotest Selection

Página 79

- 17 - 4.5. Cleaning Remember to unplug the moisture analyzer from mains before any cleaning activities. Clean the moisture analyzer using a damp c

Página 80

- 170 - MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG BALANCES AND SCALES POLAND, 26 – 600 Radom, 28 Bracka Street Phone. +

Página 81

- 18 - RS232 - socket DB9/M (male), front view: Pin2 - RxD Pin3 - TxD Pin5 - GND IN/OUT, RS232 socket DSUB15/F (female), front view: Pin1 - G

Página 82

- 19 - Usage: 1. Lack of additional sources of vibration Terminal located elsewhere than the weighing bench eliminates vibrations being a result of

Página 83

- 2 - SEPTEMBER 2014

Página 84 -  Standard printout

- 20 - Main window of a wireless connection solution features two extra icons when compared to a standard design: 1. battery charge status icon 2.

Página 85

- 21 - Terminal connected to mains, battery charged, no connection of the terminal with the weighing module. Settings For ensuring long enough ope

Página 86

- 22 - 5.2. Moisture analyzer switching off CAUTION: REMEMBER, before unplugging the moisture analyzer from mains, the terminal must be switched

Página 87

- 23 - Sending measurement result to a connected printer or computer Function key (entering instrument’s menu) Working mode selection, (programmab

Página 88

- 24 - 7. SOFTWARE STRUCTURE The structure of moisture analyzer’s main menu is divided into function groups. Each group comprises parameters groupe

Página 89

- 25 - 8. SOFTWARE MAIN WINDOW 8.9. Weighing window The software’s main window is divided into 4 main sections: • The upper part of the tou

Página 90

- 26 - • Below the workspace there is a set of soft keys: Caution: The user can declare the presence of function (quick access) keys. Declaring pr

Página 91

- 27 - • Under which there is a grey coloured workspace containing additional data on carried out activities Caution: Data contained in the worksp

Página 92

- 28 - To return to the standard means of displaying a drying result, again tap the white coloured field with visible drying result. The display retu

Página 93

- 29 - Access level Software of moisture analyzer MA 3Y series provides four access levels: administrator, advanced operator, operator and none (a

Página 94 - 19. DATABASES

- 3 - Table of contents 1. TECHNICAL SPECIFICATION ... 7 2. GENERAL SAFETY INFORMATION .

Página 95

- 30 - 10. OPERATING USER MENU MA 3Y series features a touch screen display which makes moving through software menu very simple and intuitive. Pre

Página 96

- 31 - 10.10. Return to weighing mode Changes introduced in moisture analyzer’s memory are automatically (and permanently) saved on returning to di

Página 97

- 32 - pennyweight [dwt] Taele Hongkong [tlh] Taele Singapur [tls] Taele Tajwan [tlt] Taele Chiny [tlc] Momme [mom] Grain [gr] Newton [N

Página 98

- 33 - 11.3. Leveling A moisture analyzer MA 3Y series features an electronic level control system – AutoLEVEL System: The AutoLEVEL System in

Página 99

- 34 - 11.4. Zeroing In order to zero mass indication press key. The mass indication on the display should equal zero, and precise zero and sta

Página 100 - Caution:

- 35 - PROCEDURE 1 - on taking the tarred load off moisture analyzer’s weighing pan • Press ZERO key • The NET marker is deleted, and new zero p

Página 101

- 36 - • SINGLE, mass value stored in instrument’s memory on single pressing of TARE key, the following pressing of TARE key determines new value of

Página 102 - - 102

- 37 - Never – printout deactivated First stable – the first stable measurement result obtained after placing a load on moisture analyze

Página 103

- 38 - An authorized RADWAG representative determines the minimum sample dependent on packaging mass. The process is carried out with mass standards,

Página 104

- 39 - threshold, then the <Minimum sample> mode has to be enabled in the additional setting of the weighing mode. Procedure: 1. Press grey

Página 105

- 4 - 11.2. Means of correct weighing procedure ... 32 11.3. Leveling ...

Página 106

- 40 - - Mass below the set value of minimum sample - Mass above the set value of minimum sample for a given tare range Meaning of the pictograms

Página 107 - Printout template

- 41 - Caution: If more than one reference tare value is programmed (with corresponding minimum sample mass) then the indicated value automatically m

Página 108

- 42 - 12. ADJUSTMENT Moisture analyzers MA 3Y series feature adjustment (calibration) option, which ensures their correct measurement accuracy.

Página 109 - - 109

- 43 - • On completing the adjustment procedure the moisture analyzer indicates a message box: • On confirming the message box with key, the m

Página 110

- 44 - chamber of a moisture analyzer. Before starting temperature adjustment process, place the temperature calibration kit inside the drying chambe

Página 111 - - 111

- 45 - Before starting the temperature adjustment process, close drying chamber lid and follow below process descriptions: Enter submenu group

Página 112 - - 112

- 46 - After 8 minutes, the system opens a window with an on-screen numeric keyboard. Using the on-screen keyboard insert the current temperature va

Página 113 - 20. COMMUNICATION

- 47 - 12.4. Drying chamber test Parameter <Drying chamber test> enables monitoring the temperature as indicated by the moisture analyzer’s in

Página 114

- 48 - After test completion (i.e. after testing time) the software displays window with an on-screen keyboard for entering the temperature value a

Página 115 - - 115

- 49 - An example of a report from drying chamber test: --------- Drying chamber test --------- Start time 2012.10.30 13:

Página 116 - - 116

- 5 - 18.3. Printouts in drying mode ... 83 18.4. Report from completed drying pr

Página 117 - 21. PERIPHERAL DEVICES

- 50 - 12.5. Adjustment report printout Parameter < Report printout > refers to mass adjustment process only and it determines whether or not

Página 118 - - 118

- 51 - If internal procedures of an organization require maintaining complete documentation from all carried out adjustment processes, then the list

Página 119 - Caution:

- 52 - 13. USERS Menu “Users” contains list of operators, who are authorized to operate the moisture analyzer. The following data can be defined f

Página 120

- 53 - 14. ACCESS LEVEL Access level in a moisture analyzer determines scope of activities that a user can carry out. This menu can only be modi

Página 121

- 54 - Databases The Administrator is also authorized to set access level required to modifying each of the databases. Procedure: Enter group of

Página 122 - 22. INPUTS / OUTPUTS

- 55 - 15. PROFILES A Profile is a data pack determining: o functioning of working modes, e.g. weighing and drying, o type of data displayed

Página 123 - - 123

- 56 - Caution: Adding a profile is only enabled after logging in as an Administrator . Deleting a profile Procedure: • Enter moisture anal

Página 124 - 23. OTHER PARAMETERS

- 57 - 15.3. Settings Name On entering this option, the display opens a message box with keyboard. Insert name of a profile and accept it by pres

Página 125

- 58 - FILTER Each measurement signal before being displayed is electronically processed for the purpose of obtaining corrected parameters specific

Página 126

- 59 - Last digit The function determines visibility of the last decimal place indicated on measurement result. The functions provides three settin

Página 127 - - 127

- 6 - 23.13. Autotest ...128 23.14. Start logo ...

Página 128 - - 128

- 60 - 16. WORKING MODES – general information A moisture analyzer MA 3Y series in its standard version features the following working modes: We

Página 129 - 24. UPDATE

- 61 - o Select name of a desired working mode, and press it 16.2. Parameters related to a working mode Each working mode has programmable param

Página 130 - 25. SYSTEM INFORMATION

- 62 - 16.3.1. Proximity sensors The balance comprises two proximity sensors enabling wave control of the balance therefore the user does not have t

Página 131 - 26. MOVIES

- 63 - selected, then the variable is disabled, i.e. it is not printed. PROCEDURE: 1. Press field with name of a template to be edited (Header – Wei

Página 132 - - 132

- 64 - − User − Level status − Client − Warehouse − Product − Packaging − Universal variable 1 … 5 − Dashes − Empty line − GLP report − Si

Página 133 - 27. COMMUNICATION PROTOCOL

- 65 - the software on printout). Each non-standard printout is a separate project, featuring specific name, by which it is identified, and saved in

Página 134

- 66 - 17.1.1. Means of moisture content measurement on a moisture analyzer Measurement of moisture content in a tested sample is carried out by de

Página 135 - - 135

- 67 - uniformity and other. The sampling process should be carried out as quickly as possible, so that a sample does not loose or absorb humidity fr

Página 136

- 68 - The surface of such substances often may caramelize during testing procedure. Therefore, it is recommended to use thin layer of a sample, and

Página 137 - - 137

- 69 - Obtaining short measurement time is feasible in case of small sample mass, but sample mass cannot be too small, as it may negatively influence

Página 138 - - 138

- 7 - 1. TECHNICAL SPECIFICATION Type MA 60.3Y MA 200.3Y Reading unit 0.1 mg 1 mg Max. capacity 60 g 200 g Moisture reading accuracy 0,0001% 0,00

Página 139 - - 139

- 70 - • Compare result obtained on a moisture analyzer with the one of traditional method When drying a sample with high moisture content, it is

Página 140 - - 140

- 71 - place at full power of the filaments, causing heating the drying chamber to increased temperature (maximum temperature value is higher by 30 %

Página 141 - - 141

- 72 - Automatic finish mode – user defined criterion (change of moisture content) An operator determines limit loss of moisture content over time

Página 142 - - 142

- 73 - 1. Setup – access to moisture analyzer’s menu 2. Database of drying programs – selection of drying programs 3. Database of products – se

Página 143 - - 143

- 74 - 2 This field enables setting finish mode criterion, and its parameters (depending on selected option). 3 Field enables setting type of data p

Página 144 - - 144

- 75 - After selecting desired mode, accept the choice. The software returns to displaying the previous window, and displayed data refer to selecte

Página 145 - ends

- 76 - ATTENTION: For the fast and step mode the drying time in overdrive temperature (step mode) or particular thresholds (fast mode) should be set.

Página 146 - 30. ERROR MESSAGES

- 77 - The software returns to displaying the previous window, and displayed data refers to selected finish mode. Set parameters to user needs by pre

Página 147 - - 147

- 78 - PRINTOUT Select and set parameters of printout and printing interval by pressing dedicated soft key on moisture analyzer’s display. After pr

Página 148 - - 148

- 79 - 18.2.2. Use of menu <Settings> After entering the moisture analyzer settings, the user can define all drying parameters, and main da

Página 149 - - 149

- 8 - 2. GENERAL SAFETY INFORMATION 2.1. Definitions of signals and warning signs Safety precautions are marked with special descriptions and wa

Página 150 - - 150

- 80 - Mandatory - option enabled, the display shows a bar graph demonstrating graphic visualization of sample mass in relation to mass value saved

Página 151 - 32. ADDITIONAL EQUIPMENT

- 81 - • Assign <Checkweighing thresholds> option to one of the quick access keys displayed in the bottom bar • Return to drying mode • Pres

Página 152 - - 152

- 82 - Caution: The checkweighing thresholds and target mass options may function simultaneously during drying process. In such case, the toleranc

Página 153 - - 153

- 83 - The software automatically adds a new entry to the database and immediately enters its editing mode. Fill in all mandatory parameters on a new

Página 154 - - 154

- 84 -  Standard printout comprises three internal sections featuring different variables. For each variable set option YES to include it in a stan

Página 155 - - 155

- 85 - − Unit − Printout interval − Universal variable 1 … 5 − Initial mass − Empty line − GLP report − Non-standard printout o WEIGHING −

Página 156 - - 156

- 86 - 18.4. Report from completed drying process On completing each drying process a report is generated. The report is saved in the database <

Página 157 - - 157

- 87 - 18.5. Means of carrying out drying process Bottom section of moisture analyzer’s main display features three pictograms, which indicate cur

Página 158 - - 158

- 88 - If drying parameters are correct, a user can proceed to testing process. Press key to start drying process. The software of a moisture a

Página 159 - - 159

- 89 - The display indicates initial mass of tested sample. In order to ensure stable measurement conditions, it is recommended to close the drying

Página 160 - - 160

- 9 - 2.2. Safety information CAUTION! The use of moisture analyzer MA 3Y regardless of both, safety information and user manual guidelines may caus

Página 161

- 90 - On stabilization of measurement result (indicated by stability pictogram on terminal’s display) close the weighing chamber’s lid. The sof

Página 162 - - 162

- 91 - During drying process, the display shows instantaneous measurement results and data on the testing process (testing time, temperature in the

Página 163 - - 163

- 92 - In order to finish the process press button. The moisture analyzer returns to displaying the main window. After removing the dried sample f

Página 164 - - 164

- 93 - In order to return to the main window of drying mode, press < Chart > soft key for the another time. Previewing a chart in the databa

Página 165 - - 165

- 94 - The bottom bar contains the following soft keys (options): Return to displaying the complete chart Reducing the screen to the previous view

Página 166 - - 166

- 95 - WAREHOUSES PRINTOUTS UNIVERSAL VARIABLES DATABASES MANAGEMENT 19.1. Processes carried out on databases Only authorized operator

Página 167 - - 167

- 96 - • EKSPORT – allows to export data from a database to a memory stick. Plug the memory stick to an USB port to let the program detect it and st

Página 168 - - 168

- 97 - Database of products contains names of all products that are weighed or dried. Procedure • Enter submenu < Databases>, and press <

Página 169 - Autotest Selection

- 98 - • Enter database of < Weighing records> and press desired entry (record). List of parameters on a weighing record saved in the data

Página 170 - RADWAG BALANCES AND SCALES

- 99 - 8. Printout [type of printout, label declared to a client] 19.5. Drying program Database of drying programs contains saved data o

Comentários a estes Manuais

Sem comentários