18. ERROR MESSAGES
If moisture balance software detects an error a proper error message appears on
the graphic display. The error message contains number of error which informs
about its source.
Error message
Error description
Er1 Hi
Start mass exceeds permissible range (it refers to start
and adjustmentstart mass)
Er2 nu
Low limit of converter range exceeded
Er3 Fu
High limit of converter range exceeded
Er4 ro
Value of mass or temperature exceeds permissible range
(it refers to mass or temperature adjustmentfinal mass
or temperature)
Er5 Ad
A/D converter error
Er6 to
Operation time exceeded (tarring or printing caused by
lack of stability status in set time)
Er7 ou
Inserted value is out of range (reg. setting reference
mass, date, etc.)
19. PRINCIPLES OF USE
For measurement temperatures ranging from 161°C to 250°C time interval for
temperature meintenance is estimated proportionally as follows, 60 min for 161°C
20 min for 250°C.
For drying process in temperature 250°C, Max temp is maintained for 20 min, next
the program automatically lowers the temperature to 161°C, without aborting the
drying process.
Time interval for lowering the temperature to 160°C is 20 min.
For drying mode Quick overdriving Max drying temperature is 30% but no more
than Max default temperature.
For drying mode Step time interval for temperature maintenance is limited to 20
minutes for particular steps.
PRIOR VERSION OF MOISTURE ANALYZER
Parameter <P6-06Preuer> must be set to value <YES>.