In the overlap <Transmission setting>:
Set the communication type „Ethernet Transmission TCP”,
„IP address 1” the same as <P2.4.2 IP ADDRESS> in the scale,
Port”/”Com” the same as <P2.4.5 LOCAL PORT> in the scale,
Confirm the changes by clicking ,
Rerun the program,
Press the „Connect” button,
The appropriate communication with scales is signalled by pictogram
in the left part of the display,
Transmission problems are signalled by the message window:
32.3. Example of dosing application
A standard WPW scale (3 inputs / 3 outputs in connector)
have a task to perform a two-threshold dosing (bulk/precise dosing)
in the manual mode with external buttons START and STOP process.
Caution:
Simple dosing systems can use one of three different methods: 1 – manual
mode, 2 – automatic mode, 3 – negative dosing (see ch.. 24.8.1).
Procedure:
1. According to ch.11.2.:
Enter the group of parameters <P8.1 EXTERNAL BUTTONS>
and configure the buttons according to the table below:
TARE B. NO
PRINT B. NO
ZERO B. NO
START B. IN1
STOP B. IN2
EXTERN. START NO
- 156 -