Radwag MA 200.3Y Moisture Analyzer Manual do Utilizador Página 135

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 170
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 134
- 135 -
27.2. Responses format for commands sent from computer level
On receipt of a command, the terminal sends a response in one of the
following formats:
XX_A CR LF command understood and in progress
XX_D CR LF
command carried out (appears only after the
command XX_A)
XX_I CR LF
command understood but not accessible at this
moment
XX _ ^ CR LF command understood but max range is exceeded
XX _ v CR LF command understood but min range is exceeded
XX _ OK CR LF Command carried out (completed)
ES_CR LF Command not recognized
XX _ E CR LF
an error occurred on command carrying out (time
limit exceeded while waiting for stable measurement
result (time limit is moisture analyzer’s characteristic
parameter)
XX - stands for a name of sent command
_ - substitutes spaces
COMMANDS DESCRIPTION
CAUTION
Spaces provided in commands format have to be skipped, they are
provided for improving text readability only.
Zero the moisture analyzer
Format: Z CR LF
Accessible responses:
Z_A CR LF
Z_D CR LF
- command understood and in progress
- command carried out
Z_A CR LF
Z_^ CR LF
- command understood and in progress
- command understood but zeroing range exceeded
Z_A CR LF
Z_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable measurement
result
Z_I CR LF
- command understood but not accessible at this moment
Vista de página 134
1 2 ... 130 131 132 133 134 135 136 137 138 139 140 ... 169 170

Comentários a estes Manuais

Sem comentários